Recent developments in radionuclide neutron source emission rate measurements at the National Physical Laboratory

This paper outlines the key features of the new manganese bath facility of the National Physical Laboratory and describes the measurements performed to characterise and validate the facility. Other developments have focused on the use of MCNP to calculate the fraction of neutrons captured by nuclei...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied radiation and isotopes 2010-04, Vol.68 (4), p.626-630
Hauptverfasser: Roberts, N.J., Jones, L.N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper outlines the key features of the new manganese bath facility of the National Physical Laboratory and describes the measurements performed to characterise and validate the facility. Other developments have focused on the use of MCNP to calculate the fraction of neutrons captured by nuclei other than manganese and the fraction escaping from the bath. This has highlighted the large difference in the oxygen(n,α) capture fraction when different cross-section evaluations are applied. This can change the value obtained for the neutron emission rate of a source by up to 0.8%.
ISSN:0969-8043
1872-9800
DOI:10.1016/j.apradiso.2009.09.023