Novel detection of aerosols: combined cavity ring-down and fluorescence spectroscopy
High fluences inside cavity ring-down spectroscopy optical resonators lend themselves to fluorescence or Raman spectroscopy. An instrument at 488 nm was developed to measure extinction, and fluorescence of aerosols. A detection limit of 6 x 10;-9 cm;-1Hz;-1/2 (0.6 Mm;-1Hz;-1/2) was achieved. The flu...
Gespeichert in:
Veröffentlicht in: | Optics express 2005-05, Vol.13 (9), p.3376-3387 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High fluences inside cavity ring-down spectroscopy optical resonators lend themselves to fluorescence or Raman spectroscopy. An instrument at 488 nm was developed to measure extinction, and fluorescence of aerosols. A detection limit of 6 x 10;-9 cm;-1Hz;-1/2 (0.6 Mm;-1Hz;-1/2) was achieved. The fluorescence spectral power collected from a single fluorescent microsphere was 10 to 20 pW/nm. This power is sufficient to obtain the spectrum of a single microsphere with a resolution of 10 nm and signal-to-noise ratio of ~10. The relative concentrations of two types of fluorescent microspheres were determined from a time-integrated fluorescence measurement of a mixture of both. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/opex.13.003376 |