Resolution of Heart Block After Surgical Removal of an Amplatzer Device

A large secundum atrial septal defect in a 4-year-old child was closed by percutaneous placement of an Amplatzer septal occluder (AGA Medical Corporation, Plymouth, MN). After device placement, complete heart block developed that did not resolve after 3 days of medical management. The patient subseq...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 2010-05, Vol.89 (5), p.1631-1633
Hauptverfasser: Clark, Joseph B., MD, Chowdhury, Devyani, MD, Pauliks, Linda B., MD, Weber, Howard S., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A large secundum atrial septal defect in a 4-year-old child was closed by percutaneous placement of an Amplatzer septal occluder (AGA Medical Corporation, Plymouth, MN). After device placement, complete heart block developed that did not resolve after 3 days of medical management. The patient subsequently underwent surgical removal of the device and suture closure of the atrial septal defect. The patient recovered conduction after the operation. We report the successful treatment of device-induced heart block with surgical removal of the device.