Report of the IFCC Working Group for Standardization of Thyroid Function Tests; Part 1: Thyroid-Stimulating Hormone
Laboratory testing of serum thyroid-stimulating hormone (TSH) is an essential tool for the diagnosis and management of various thyroid disorders whose collective prevalence lies between 4% and 8%. However, between-assay discrepancies in TSH results limit the application of clinical practice guidelin...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2010-06, Vol.56 (6), p.902-911 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Laboratory testing of serum thyroid-stimulating hormone (TSH) is an essential tool for the diagnosis and management of various thyroid disorders whose collective prevalence lies between 4% and 8%. However, between-assay discrepancies in TSH results limit the application of clinical practice guidelines.
We performed a method comparison study with 40 sera to assess the result comparability and performance attributes of 16 immunoassays.
Thirteen of 16 assays gave mean results within 10% of the overall mean. The difference between the most extreme means was 39%. Assay-specific biases could be eliminated by recalibration to the overall mean. After recalibration of singlicate results, all assays showed results within the biological total error goal (22.8%), except for 1 result in each of 4 assays. For a sample with a TSH concentration of 0.016 mIU/L, 6 assays either did not report results or demonstrated CVs >20%. Within-run and total imprecision ranged from 1.5% to 5.5% and 2.5% to 7.7%, respectively. Most assays were able to match the internal QC targets within 5%. Within-run drifts and shifts were observed.
Harmonization of TSH measurements would be particularly beneficial for 3 of the 16 examined assays. These data demonstrate that harmonization may be accomplished by establishing calibration traceability to the overall mean values for a panel of patient samples. However, the full impact of the approach must be further explored with a wider range of samples. Although a majority of assays showed excellent quality of performance, some would benefit from improved within-run stability. |
---|---|