Clinical comparison of automated and manual keratometry in pre-operative ocular biometry

Corneal measurements, using the manual (Topcon OM-4) and automated (Canon RK-1) keratometers was performed on 104 eyes of 104 patients undergoing cataract and implant surgery to assess the role of automated keratometry in pre-operative ocular biometry. Four eyes of four patients were excluded from s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Eye (London) 1992-01, Vol.6 (1), p.60-62
1. Verfasser: SunderRaj, Palaniswamy
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Corneal measurements, using the manual (Topcon OM-4) and automated (Canon RK-1) keratometers was performed on 104 eyes of 104 patients undergoing cataract and implant surgery to assess the role of automated keratometry in pre-operative ocular biometry. Four eyes of four patients were excluded from statistical analyses for various reasons. The time taken to perform automated keratometry was a mean of 61 (SD 21) seconds compared to 205 (SD 37) seconds for manual keratometry; the difference was statistically significant (p