A comparison of a second- and a third-generation modular cup design: is new improved?
The highly cross-linked polyethylene liners currently used with modular uncemented cups have substantially decreased wear and osteolysis at early follow-up. However, retroacetabular osteolysis has still been reported in some cases with DePuy Orthopaedic's (Warsaw, IN) second-generation Duraloc...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2010-06, Vol.25 (4), p.514-521 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The highly cross-linked polyethylene liners currently used with modular uncemented cups have substantially decreased wear and osteolysis at early follow-up. However, retroacetabular osteolysis has still been reported in some cases with DePuy Orthopaedic's (Warsaw, IN) second-generation Duraloc acetabular shell. DePuy's third-generation Pinnacle cup incorporates a different shell-liner locking mechanism. We compared the clinical outcome among a matched series of 42 Duraloc and 42 Pinnacle cups at a mean follow-up of 5.9 years. Although the Harris Hip Scores and wear rates were not statistically different between the 2 cup designs, retroacetabular osteolysis behind the central hole was absent among the Pinnacle cups but noted among 19% of the Duraloc cups. |
---|---|
ISSN: | 1532-8406 |
DOI: | 10.1016/j.arth.2009.02.018 |