The three-dimensional bone interface of an osseointegrated implant: A method for study

To clarify the three-dimensional bone structure around a plasma-sprayed hydroxyapatite-coated endosseous dental implant, bone contact with the implant at 75 μm intervals was evaluated. Based on digitized data, three-dimensional graphics of the bone around the implant was clearly shown. These graphic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of prosthetic dentistry 1992-11, Vol.68 (5), p.813-816
Hauptverfasser: Akagawa, Yasumasa, Wadamoto, Masayoshi, Sato, Yuuji, Tsuru, Hiromichi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To clarify the three-dimensional bone structure around a plasma-sprayed hydroxyapatite-coated endosseous dental implant, bone contact with the implant at 75 μm intervals was evaluated. Based on digitized data, three-dimensional graphics of the bone around the implant was clearly shown. These graphics provide adequate information on the bone-implant interface and also suggest development of a realistic finite element model.
ISSN:0022-3913
1097-6841
DOI:10.1016/0022-3913(92)90208-R