High-definition freehand 3-D ultrasound

This paper describes a high-definition freehand 3-D ultrasound (US) system, with accuracy surpassing that of previously documented systems. 3-D point location accuracy within a US data set can be achieved to within 0.5 mm. Such accuracy is possible through a series of novel system-design and calibra...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 2003-04, Vol.29 (4), p.529-546
Hauptverfasser: Treece, Graham M, Gee, Andrew H, Prager, Richard W, Cash, Charlotte J.C, Berman, Laurence H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper describes a high-definition freehand 3-D ultrasound (US) system, with accuracy surpassing that of previously documented systems. 3-D point location accuracy within a US data set can be achieved to within 0.5 mm. Such accuracy is possible through a series of novel system-design and calibration techniques. The accuracy is quantified using a purpose-built tissue-mimicking phantom, designed to create realistic clinical conditions without compromising the accuracy of the measurement procedure. The paper includes a thorough discussion of the various ways of measuring system accuracy and their relative merits; and compares, in this context, all recently documented freehand 3-D US systems. (E-mail: gmt11@eng.cam.ac.uk)
ISSN:0301-5629
1879-291X