Design and evaluation of shielded gradient coils
Techniques are described for the design of shielded gradient coils for superconducting MRI systems. These design methods are suited for constructing the most efficient gradient coil that meets a specified homogeneity requirement. Tradeoffs in coil design of efficiency with coil size and gap size are...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 1992-08, Vol.26 (2), p.191-206 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Techniques are described for the design of shielded gradient coils for superconducting MRI systems. These design methods are suited for constructing the most efficient gradient coil that meets a specified homogeneity requirement. Tradeoffs in coil design of efficiency with coil size and gap size are discussed. Residual eddy currents from coils constructed with a finite number of wires are calculated and give guidelines for the construction of efficient, whole‐body gradient coils. |
---|---|
ISSN: | 0740-3194 1522-2594 |
DOI: | 10.1002/mrm.1910260202 |