A method for pneumatically inserting an array of penetrating electrodes into cortical tissue

The goal of this research was to find a practical means by which an array of 100 needle-shaped electrodes could be implanted into the cerebral cortex with minimal brain tissue trauma. It was found that insertion of these structures into cortical tissues could only be performed using high insertion s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of biomedical engineering 1992-01, Vol.20 (4), p.413-422
Hauptverfasser: ROUSCHE, P. J, NORMANN, R. A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The goal of this research was to find a practical means by which an array of 100 needle-shaped electrodes could be implanted into the cerebral cortex with minimal brain tissue trauma. It was found that insertion of these structures into cortical tissues could only be performed using high insertion speeds. A pneumatically actuated impact insertion system has been developed that is capable of inserting an electrode array into feline brain tissue at speeds from about 1 to 11 m/s. We found that a minimum array insertion speed of 8.3 m/s was necessary for a complete, safe insertion of all 100 electrodes in the array to a depth of 1.5 mm into feline cortex. The performance of the impact insertion system is discussed in terms of a simplified representation of cortical tissue.
ISSN:0090-6964
1573-9686
DOI:10.1007/BF02368133