Continuous thermodilution cardiac output measurement in sheep

A technique has been developed to continuously measure cardiac output by means of the principles of thermodilution. Pulmonary artery catheters were modified by placing a 10 cm filament near the usual injectate port. Small amounts of heat were infused according to a randomly repeating binary on-off s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of thoracic and cardiovascular surgery 1992-08, Vol.104 (2), p.315-320
Hauptverfasser: Yelderman, Mark, Quinn, Michael D., McKown, Russell C., Eberhart, Robert C., Dollar, Michael L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A technique has been developed to continuously measure cardiac output by means of the principles of thermodilution. Pulmonary artery catheters were modified by placing a 10 cm filament near the usual injectate port. Small amounts of heat were infused according to a randomly repeating binary on-off sequence. The distal blood temperature was recorded and cross-correlated with the heat waveform to produce a dilution curve and calculate cardiac output. The technique was compared with bolus thermodilution in seven sheep. Cardiac output ranged from 1.5 to 13.2 L/min, and heart rate varied from 59 to 180 beats/min. The linear regression between the data obtained by the two methods is represented by the equation y = 1.00x + 0.13; the correlation coefficient, R, is 0.97, and the p value is
ISSN:0022-5223
1097-685X
DOI:10.1016/S0022-5223(19)34783-X