Experimental Studies of Electroosmotic Flow Dynamics in Microfabricated Devices during Current Monitoring Experiments

Electroosmotic flow (EOF) was monitored in glass microfluidic devices at rates up to 2 Hz with a precision of 0.2−1.0% using a technique based on the periodic photobleaching of a dilute, neutral fluorophore added to the running buffer. This EOF monitoring method was used to examine the performance o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2003-02, Vol.75 (3), p.361-370
Hauptverfasser: Pittman, Jason L, Henry, Charles S, Gilman, S. Douglass
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Electroosmotic flow (EOF) was monitored in glass microfluidic devices at rates up to 2 Hz with a precision of 0.2−1.0% using a technique based on the periodic photobleaching of a dilute, neutral fluorophore added to the running buffer. This EOF monitoring method was used to examine the performance of the current monitoring technique for measuring an average electroosmotic flow in a microfluidic device with a cross-T design. Flow measurements made with the current monitoring method gave a precision of 0.4−2.2%, but the periodic photobleaching method shows that the current monitoring technique causes changes in EOF as high as 41% during a single experiment. The periodic photobleaching method for EOF monitoring was also used to study EOF in channels on opposite sides of a cross-channel intersection. The opposite channels were shown to exhibit substantially different EOF dynamics during a current monitoring experiment as well as different steady-state EOF rates during normal operating conditions.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac026132n