Three-dimensional distribution of radiation within the breast: an intercomparison of departments participating in the START trial of breast radiotherapy fractionation
Purpose : To examine the ability of computer planning systems to calculate the dose to the breast correctly in three dimensions. Both the absolute dose at the center of the breast and the accuracy of the isodose distributions were investigated. Methods and Materials : Measurements were performed in...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 2003, Vol.55 (1), p.271-279 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
: To examine the ability of computer planning systems to calculate the dose to the breast correctly in three dimensions. Both the absolute dose at the center of the breast and the accuracy of the isodose distributions were investigated.
Methods and Materials
: Measurements were performed in a water-filled breast phantom using an ionization chamber. Thirty-six sets of data obtained during the Standardization of Breast Radiotherapy breast fractionation trial quality assurance program were included in the analysis. The planning systems were grouped according to the algorithms used on the basis of the definitions given in International Commission on Radiation Units and Measurements Report No. 24.
Results
: Thirty-two of the 36 planning systems overestimated the dose to the center of the breast, with a mean measured/calculated dose ratio of 0.979 (SD 0.013). The relative dose within 2 cm of the lung was also overestimated.
Conclusion
: Only one algorithm (collapsed cone) investigated in this study was able to calculate the dose at the center of the breast correctly in tangential breast radiotherapy. With modern algorithms, it is important to include a correction for the lower density of the lung, because the dose close to the interface between breast and lung tissue will also be lower than anticipated. |
---|---|
ISSN: | 0360-3016 1879-355X |
DOI: | 10.1016/S0360-3016(02)03808-7 |