Chronic Threshold Testing of Implantable Cardioverter Defibrillators Does Not Increase Coagulation Activity or Platelet Activation

SMOLICH, J.J., et al.: Chronic Threshold Testing of Implantable Cardioverter Defibrillators Does Not Increase Coagulation Activity or Platelet Activation. It is unknown if the brief periods of circulatory stasis occurring with induction of VF during DFT testing in patients with an ICD activate blood...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2002-11, Vol.25 (11), p.1594-1598
Hauptverfasser: SMOLICH, JOSEPH J., YATES, LYNETTE A., MALAN, ERICA, GAN, T. ENG, HARPER, RICHARD W., ALISON, JEFFREY F., PEVERILL, ROGER E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:SMOLICH, J.J., et al.: Chronic Threshold Testing of Implantable Cardioverter Defibrillators Does Not Increase Coagulation Activity or Platelet Activation. It is unknown if the brief periods of circulatory stasis occurring with induction of VF during DFT testing in patients with an ICD activate blood coagulation processes. To address this question, coagulation activity and platelet activation were measured in peripheral venous blood samples obtained from 12 patients undergoing DFT testing under general anesthesia, 3 (n = 11) or 6 months (n = 1) after ICD implantation for recurrent ventricular arrhythmias. Five patients were anticoagulated with warfarin and two to six episodes of VF (median five) were induced per patient. Blood samples were drawn at baseline, after each DFT test and on the following morning. Coagulation activity was assessed by measuring prothrombin fragment 1+2 (F1+2) (a marker of thrombin generation), soluble fibrin (a marker of fibrin production), and D‐dimer (a breakdown product of cross‐linked fibrin). Platelet activation was evaluated by measuring the expression P‐selectin on the platelet surface using flow cytometry. No significant changes in F1+2, soluble fibrin, D‐dimer, or P‐selectin expression occurred during DFT testing, or between baseline and morning after samples. Moreover, results were unaffected by warfarin. These findings suggest that chronic threshold testing of implantable cardioverter defibrillators is not associated with activation of blood coagulation processes.
ISSN:0147-8389
1540-8159
DOI:10.1046/j.1460-9592.2002.01594.x