A venous foot pump reduces thrombosis after total hip replacement
In a prospective, randomised controlled trial, the efficacy of the A-V Impulse System in the prevention of deep-vein thrombosis was investigated in 84 patients who had undergone total hip replacement. The incidence of venographically proven, and clinically significant postoperative deep-vein thrombo...
Gespeichert in:
Veröffentlicht in: | Journal of bone and joint surgery. British volume 1992, Vol.74 (1), p.45-49 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In a prospective, randomised controlled trial, the efficacy of the A-V Impulse System in the prevention of deep-vein thrombosis was investigated in 84 patients who had undergone total hip replacement. The incidence of venographically proven, and clinically significant postoperative deep-vein thrombosis was 40% in the control group and 5% in the treatment group (p less than 0.001). No adverse reactions were recorded. |
---|---|
ISSN: | 0301-620X 2044-5377 |
DOI: | 10.1302/0301-620x.74b1.1732264 |