Diagnostic ultrasound artifacts during imaging of two-body interfaces: Part 2––beam thickness artifact

A diagnostic ultrasound method is being developed for measuring surface contact areas at the tibio–femoral interface of a total knee replacement in a non-clinical industrial setting as an engineering design tool. As an initial step towards this, a previous study mathematically predicted the effect o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasonics 2002-10, Vol.39 (10), p.689-693
Hauptverfasser: Zdero, R, Fenton, P.V, Bryant, J.T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A diagnostic ultrasound method is being developed for measuring surface contact areas at the tibio–femoral interface of a total knee replacement in a non-clinical industrial setting as an engineering design tool. As an initial step towards this, a previous study mathematically predicted the effect of ultrasound beam thickness on contact area measurements at a two-body interface. In the current study, a novel metal-on-polymer acoustic test object was constructed to create circular two-body interfaces of known geometry. The object was ultrasonically imaged, contact areas measured, and the results compared with the theoretical model previously developed.
ISSN:0041-624X
1874-9968
DOI:10.1016/S0041-624X(02)00382-7