Usefulness of automatic detection of fragmented red cells using a hematology analyzer for diagnosis of thrombotic microangiopathy
Summary We previously reported an automatic method for quantitative analysis of schistocytes or fragmented red cells using an automatic hematology analyzer, XE‐2100. In the study reported here, we evaluated the accuracy of this detection method in patients with thrombotic microangiopathy (TMA). A fo...
Gespeichert in:
Veröffentlicht in: | Clinical and laboratory haematology 2002-12, Vol.24 (6), p.347-351 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Summary We previously reported an automatic method for quantitative analysis of schistocytes or fragmented red cells using an automatic hematology analyzer, XE‐2100. In the study reported here, we evaluated the accuracy of this detection method in patients with thrombotic microangiopathy (TMA). A follow‐up study was performed on 14 patients with two types of TMA, thrombotic thrombocytopenic purpura or hemolytic uremic syndrome. Schistocyte percent was evaluated both with an automatic counter and by means of microscopic observation. Total activity and isoenzyme pattern of lactate dehydrogenase (LD) were also determined. In these patients, schistocyte percent determined by automatic counting correlated highly with that determined by manual counting under microscopic observation (r = 0.852, P |
---|---|