The determination of platinum in biological tissue by instrumental neutron activation analysis
Neutron activation analysis was used to determine platinum in the tissues of rats that had received doses of the tumor chemotherapy agent cis-Pt. Two elements causing spectral interference were identified as 47Sc from 46Ca, a natural component of tissue, and tin ( 117mSn), an element introduced inad...
Gespeichert in:
Veröffentlicht in: | International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes Applied radiation and isotopes, 1991, Vol.42 (8), p.775-776 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Neutron activation analysis was used to determine platinum in the tissues of rats that had received doses of the tumor chemotherapy agent
cis-Pt. Two elements causing spectral interference were identified as
47Sc from
46Ca, a natural component of tissue, and tin (
117mSn), an element introduced inadvertently during necropsy and sample work-up. A protocol was devised to correct for interference, using emission lines from
47Ca and
113Sn, respectively, with appropriate consideration for relative detector efficiencies, decay times and half-lives. |
---|---|
ISSN: | 0883-2889 |
DOI: | 10.1016/0883-2889(91)90183-2 |