Comparison of the glucose oxidase method for glucose determination by manual assay and automated analyzer

In experimental models of diabetes, glucose levels in plasma and blood are commonly determined by colorimetric assay and by automated analyzers based on the glucose oxidase conversion of glucose and O 2 to gluconate and H 2O 2. We have compared the glucose levels obtained by these two methods in con...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of pharmacological and toxicological methods 2000-11, Vol.44 (3), p.543-546
Hauptverfasser: Yuen, Violet G, McNeill, John H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In experimental models of diabetes, glucose levels in plasma and blood are commonly determined by colorimetric assay and by automated analyzers based on the glucose oxidase conversion of glucose and O 2 to gluconate and H 2O 2. We have compared the glucose levels obtained by these two methods in control Wistar rats, streptozotocin diabetic Wistar rats, Zucker fa/fa fatty rats and Zucker Diabetic Fatty rats. We found that the manual glucose assay and the glucose analyzer produced comparable values up to concentrations of about 25 mM. Above this level, samples should be diluted.
ISSN:1056-8719
1873-488X
DOI:10.1016/S1056-8719(01)00117-4