Histologic analysis of a retrieved hydroxyapatite-coated femoral prosthesis

A hydroxyapatite-coated hip hemi-prosthesis was retrieved from a 98-year-old osteoporotic woman 12 weeks after implantation. Histologic analysis revealed bone and fibrous tissue almost evenly distributed around the surface of the implant circumference. Quantitative histologic analysis showed that 48...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical orthopaedics and related research 1991-11, Vol.272 (272), p.255-258
Hauptverfasser: SØBALLE, K, GOTFREDSEN, K, BROCKSTEDT-RASMUSSEN, H, NIELSEN, P. T, RECHNAGEL, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A hydroxyapatite-coated hip hemi-prosthesis was retrieved from a 98-year-old osteoporotic woman 12 weeks after implantation. Histologic analysis revealed bone and fibrous tissue almost evenly distributed around the surface of the implant circumference. Quantitative histologic analysis showed that 48% of the hydroxyapatite surface was covered by bone. Fibrous tissue covered 30% of the prosthetic surface, and 20% of the surface had no tissue coverage. Scanning electron microscopy showed direct contact without any clear boundary between the newly formed bone and the hydroxyapatite ceramic.