Single-sided membrane introduction mass spectrometry for on-line determination of semi-volatile organic compounds in air

Construction, optimization, and testing of a novel single-sided configuration for a semi-permeable [poly(dimethylsiloxane); PDMS] membrane introduction system for mass spectrometry is described. On-line detection of semi-volatile organic compounds of environmental interest is shown, including lindan...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analyst (London) 2001-11, Vol.126 (11), p.1980-1984
Hauptverfasser: RITER, Leah S, TAKATS, Zoltan, COOKS, R. Graham
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Construction, optimization, and testing of a novel single-sided configuration for a semi-permeable [poly(dimethylsiloxane); PDMS] membrane introduction system for mass spectrometry is described. On-line detection of semi-volatile organic compounds of environmental interest is shown, including lindane (a pesticide), hexahydro-1,3,5-trinitro-1,3,5-triazine (RDX) (an explosive), butylated hydroxytoluene (BHT) (an antioxidant), 1,2-dichlorobenzene, dimethylmethyl phosphonate (DMMP) (a chemical warfare agent simulant) and naphthalene. The technique has limits of detection in the sub-ppb range. with rise times of 4 to 7 s and fall times of 12 to 36 s and a response that is linear over 4 orders of magnitude (from 0.1 ppb to 1000 ppb for DMMP). The cycle time, from crude air sampling to acquisition of results, is approximately 1 min. No sample preparation is necessary.
ISSN:0003-2654
1364-5528
DOI:10.1039/b105857f