A Pilot Study of Left Tracheal Pulse Oximetry

We tested the hypothesis that left tracheal pulse oximetry (SpO(2)) is more accurate than finger SpO(2) when compared with oxygen saturation from arterial blood samples (SaO(2)) in anesthetized patients with normal thoracic anatomy. We also tested the hypothesis that tracheal oximetry readings are p...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anesthesia and analgesia 2000-10, Vol.91 (4), p.1003-1006
Hauptverfasser: Brimacombe, J., Keller, C., Margreiter, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We tested the hypothesis that left tracheal pulse oximetry (SpO(2)) is more accurate than finger SpO(2) when compared with oxygen saturation from arterial blood samples (SaO(2)) in anesthetized patients with normal thoracic anatomy. We also tested the hypothesis that tracheal oximetry readings are primarily derived from the tracheal mucosa. We studied 20 hemodynamically stable, well oxygenated, anesthetized patients with normal anatomy (ASA physical status I-III, 18-80 yr old). A single-use pediatric pulse oximeter was attached to the left lateral surface of a tracheal tube cuff. Tracheal and finger SpO(2) (dominant index finger), and SaO(2) (nondominant radial artery) were taken with the intracuff pressure at 0-60 cm H(2)O. Tracheal SpO(2) was the same as SaO(2) at an intracuff pressure of 10-60 cm H(2)O, but was less when the intracuff pressure was zero (P
ISSN:0003-2999
1526-7598
DOI:10.1097/00000539-200010000-00043