Evaluation of Two Different Homogeneous Assays for LDL-Cholesterol in Lipoprotein-X-positive Serum
The purpose of this study was to evaluate the performance of two homogeneous assays for LDL-cholesterol (LDL-C), a polyethylene/cyclodextrin (PC) assay and a detergent (D) assay, which are based on different principles, in cholestatic serum. We compared serum LDL-C concentrations determined by the t...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2000-09, Vol.46 (9), p.1351-1356 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to evaluate the performance of two homogeneous assays for LDL-cholesterol (LDL-C), a polyethylene/cyclodextrin (PC) assay and a detergent (D) assay, which are based on different principles, in cholestatic serum.
We compared serum LDL-C concentrations determined by the two assays for healthy normolipidemic subjects (n = 42) and cholestatic patients (n = 51). LDL-C concentrations obtained with the homogeneous assays were also compared with those obtained by HPLC for patients' sera. In the interference study, conjugated bile acids were added to normal serum, and their effects on the two assays were examined. The effects of lipoprotein-X (LP-X), intermediate-density lipoprotein (IDL), and apolipoprotein (apo) E-rich HDL on the LDL-C assays were also investigated by adding these lipoproteins to normal serum.
The LDL-C concentrations obtained with the D assay were higher than those obtained with the PC assay in the serum with high LP-X. The bias for LDL-C vs LP-X in cholestatic serum correlated with LP-X concentration (r = 0.582; P: |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1093/clinchem/46.9.1351 |