Elimination of Fluorescence and Scattering Backgrounds in Luminescence Lifetime Measurements Using Gated-Phase Fluorometry

A new gated form of phase fluorometry for measuring lifetimes is presented. The technique uses a square-wave excitation and gates the detector on only during the off period of the excitation. Using a long-lived sample, this eliminates or reduces errors from scattered light and short-lived fluorescen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2002-09, Vol.74 (18), p.4821-4827
Hauptverfasser: Rowe, H. M, Chan, Sing Po, Demas, J. N, DeGraff, B. A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new gated form of phase fluorometry for measuring lifetimes is presented. The technique uses a square-wave excitation and gates the detector on only during the off period of the excitation. Using a long-lived sample, this eliminates or reduces errors from scattered light and short-lived fluorescences. Using a square-wave modulated excitation source with a 50% duty cycle, traditional data treatment can be used after, at most, a simple π/2 phase adjustment. A combination of theory and experimental results demonstrates the validity of this new gated method and its utility for eliminating or reducing background. The results are precise, accurate, eliminate scattering errors, and greatly reduce errors due to short-lived fluorescence impurities. Errors from fluorescence bleed-through into the detection period or a slow excitation source turn off can be mitigated by using an offset time prior to gating the detector on.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac020150n