Lavage Solution Temperature Influences Depth of Chondrocyte Death and Surface Contouring During Thermal Chondroplasty with Temperature-Controlled Monopolar Radiofrequency Energy
Background: Although radiofrequency energy can smooth and contour cartilage surface, it has deleterious effects on chondrocyte viability. Hypothesis: Monopolar thermal chondroplasty in a 37°C lavage solution, as compared with a 22° lavage solution, will reduce chondrocyte death and result in great...
Gespeichert in:
Veröffentlicht in: | The American journal of sports medicine 2002-09, Vol.30 (5), p.667-673 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background: Although radiofrequency energy can smooth and contour cartilage surface, it has deleterious effects on chondrocyte viability.
Hypothesis: Monopolar thermal chondroplasty in a 37°C lavage solution, as compared with a 22° lavage solution, will reduce chondrocyte
death and result in greater smoothing of the articular cartilage surface.
Study Design: Controlled laboratory study.
Methods: Sixteen chondromalacic samples from patients undergoing total knee arthroplasty were divided into two groups: 22°C and 37°C
lavage solution. Each sample was divided into two equal parts and half of each group was treated for 10 seconds and the other
half for 15 seconds.
Results: Confocal laser microscopy demonstrated that the depth of chondrocyte death in the 37°C lavage solution group was significantly
less (range, 200 to 340 μm) than that in the 22°C solution group for both 10- and 15-second treatment times. Scanning electron
microscopy demonstrated that the cartilage surface in the 37°C lavage solution group was smoother than that in the 22°C solution
group for the 10-second treatment time. Energy delivery power in the 37°C lavage solution group was significantly lower than
in the 22°C solution group for both treatment times.
Conclusions: Thermal chondroplasty with 37°C lavage solution resulted in less depth of chondrocyte death and produced smoother surfaces
than with 22°C solution for 10 seconds of treatment.
Clinical Relevance: Less chondrocyte death would permit increased use of thermal chondroplasty. |
---|---|
ISSN: | 0363-5465 1552-3365 |
DOI: | 10.1177/03635465020300050601 |