Type III Endoleak from a Thoracic Aortic Stent-Graft
Purpose: To present an as yet unreported late complication of an Excluder thoracic endograft. Case Report: A 78-year-old man underwent surgery for a ruptured type V thoracoabdominal aortic aneurysm in 1996. Four years later, an aneurysm was detected in the proximal thoracic aorta and repaired with 2...
Gespeichert in:
Veröffentlicht in: | Journal of endovascular therapy 2002-08, Vol.9 (4), p.535-538 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
To present an as yet unreported late complication of an Excluder thoracic endograft.
Case Report:
A 78-year-old man underwent surgery for a ruptured type V thoracoabdominal aortic aneurysm in 1996. Four years later, an aneurysm was detected in the proximal thoracic aorta and repaired with 2 Excluder endoprostheses. At 12 months, computed tomography showed an increase in the aneurysm sac diameter and a type III endoleak, which was traced to a hole in the stent-graft fabric on arteriography. No fracture of the metal components was detected in the stent-grafts. Another Excluder device was implanted within the distal endograft. Satisfactory exclusion of the leak has been maintained for 6 months.
Conclusions:
The risk of type III leaks must be minimized before stent-grafting can be regarded as a routine procedure in the treatment of thoracic aortic aneurysms. |
---|---|
ISSN: | 1526-6028 1545-1550 |