Intravascular Drug Delivery With a Pulsed Liquid Microjet

Occlusions of the retinal veins and arteries, associated with diseases such as hypertension and arteriosclerosis, are a major cause of severe and irreversible loss of vision. Treatments for retinal vascular diseases have been unsatisfactory owing in part to the difficulty of delivering drugs to the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Archives of ophthalmology (1960) 2002-09, Vol.120 (9), p.1206-1208
Hauptverfasser: Fletcher, Daniel A, Palanker, Daniel V, Huie, Philip, Miller, Jason, Marmor, Michael F, Blumenkranz, Mark S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Occlusions of the retinal veins and arteries, associated with diseases such as hypertension and arteriosclerosis, are a major cause of severe and irreversible loss of vision. Treatments for retinal vascular diseases have been unsatisfactory owing in part to the difficulty of delivering drugs to the site of disease within the eye. In this article, we demonstrate that a new device, the vapor bubble–driven pulsed liquid microjet, can deliver drugs into the lumen of small vessels such as those found in the retina. A 15-µm-diameter liquid jet traveling at more than 60 m/s was shown to penetrate and deliver fluid through the wall of a blood vessel that was 60µm in diameter. Perforation of the wall of the blood vessel did not extend beyond the jet diameter.Arch Ophthalmol.2002;120:1206-1208-->