Detection Limit of Isotope Dilution Mass Spectrometry

The detection limit is an important figure of merit for evaluating instrumentation and analytical methods. While the detection limit for techniques using linear calibration functions has been studied extensively, this fundamental metric has rarely been discussed for mass spectrometry that bases the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2002-08, Vol.74 (15), p.3887-3891
Hauptverfasser: Yu, Lee L, Fassett, John D, Guthrie, William F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The detection limit is an important figure of merit for evaluating instrumentation and analytical methods. While the detection limit for techniques using linear calibration functions has been studied extensively, this fundamental metric has rarely been discussed for mass spectrometry that bases the calibration on the principle of isotope dilution. We have developed a formulation for the detection limit for isotope dilution mass spectrometry (IDMS) after a thorough analysis of the uncertainty of IDMS measurements. The new formulation describes the IDMS detection limit as a function of the enrichment of the isotopic spike and the linear calibration detection limits measured at the masses for the isotope ratio measurement.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac011254l