The influence of environmental parameters in electrostatic cell radon monitor response
To better understand the influence of air parameters such as pressure, humidity and temperature on the counting efficiency of a radon monitor based on solid state silicon detectors and radon daughter electrostatic collection, a number of experimental and theoretical studies have been performed. This...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2004-08, Vol.61 (2), p.243-247 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To better understand the influence of air parameters such as pressure, humidity and temperature on the counting efficiency of a radon monitor based on solid state silicon detectors and radon daughter electrostatic collection, a number of experimental and theoretical studies have been performed. This study has been carried out using a multiparameter acquisition system used to monitor radon, temperature, pressure and relative humidity. Results show that air humidity and temperature inside the cell influence radon daughter collection in a significant way. For the tested cell, a decrease of the
218Po counting efficiency vs. humidity and temperature increase was measured in the ranges from 10% to 90% and from 15°C to 35°C, respectively. These effects were also theoretically studied by using Monte-Carlo software, which takes into account electrostatic collection of radon daughters in the cell, by considering the recombination process as a function of climatic parameters. A satisfactory agreement was obtained between experimental measurements and theoretical calculations. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2004.03.053 |