Observation of significant signal voids in images of large biological samples at 11.1 T

Proton MRI of large biological samples were obtained on an 11.1 T / 40 cm instrument. Images were obtained of a fixed human brain and a large piece of fresh beef. The proton MR images demonstrate severe distortions within these conductive samples, indicative of shortened electrical wavelengths and w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2004-06, Vol.51 (6), p.1103-1107
Hauptverfasser: Beck, Barbara L., Jenkins, Kelly, Caserta, Jim, Padgett, Kyle, Fitzsimmons, Jeffrey, Blackband, S.J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Proton MRI of large biological samples were obtained on an 11.1 T / 40 cm instrument. Images were obtained of a fixed human brain and a large piece of fresh beef. The proton MR images demonstrate severe distortions within these conductive samples, indicative of shortened electrical wavelengths and wave behavior within the sample. These observations have significant implications with respect to the continuing evolution of MR to higher magnetic field strengths on large samples, particularly on humans. Magn Reson Med 51:1103–1107, 2004. © 2004 Wiley‐Liss, Inc.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.20120