Recent advances in ion chromatography suppressor improve anion separation and detection
The Alltech DS-Plus suppressor improves and simplifies anion analysis by suppressor-based ion chromatography (IC). In addition to suppressing the mobile phase and enhancing the analyte signal, the DS-Plus suppressor removes carbonic acid from the suppressor effluent when carbonate–hydrogencarbonate...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 2002-05, Vol.956 (1), p.15-22 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Alltech DS-Plus suppressor improves and simplifies anion analysis by suppressor-based ion chromatography (IC). In addition to suppressing the mobile phase and enhancing the analyte signal, the DS-Plus suppressor removes carbonic acid from the suppressor effluent when carbonate–hydrogencarbonate mobile phases are used, reducing the water dip that interferes with early eluting peaks, increasing detection sensitivity, and enabling carbonate–hydrogencarbonate gradients. The ability to operate with carbonate–hydrogencarbonate gradients places exceptional separating power in the mobile phase, forgoing the need for expensive columns with special selectivity. Continuous operation and solid-phase chemistry eliminates the need for external regenerants, switching valves, and multiple operating modes. |
---|---|
ISSN: | 0021-9673 |
DOI: | 10.1016/S0021-9673(01)01515-1 |