Durable Microfabricated High-Speed Humidity Sensors
We describe a durable microfabricated humidity sensor made of interdigitated rhodium electrodes on a silicon substrate covered with a sensing film of Nafion perfluorosulfonate ionomer. Rhodium electrodes are much less prone to oxidative degradation compared to previously described gold electrode-bas...
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2004-05, Vol.76 (9), p.2561-2567 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe a durable microfabricated humidity sensor made of interdigitated rhodium electrodes on a silicon substrate covered with a sensing film of Nafion perfluorosulfonate ionomer. Rhodium electrodes are much less prone to oxidative degradation compared to previously described gold electrode-based sensors. Even with dc excitation, Rh electrode sensors exhibit excellent long-term response stability. It has been found that low-amplitude (±1 V) square wave excitation can prolong the usability of gold electrode-based sensors to at least several months; however, this mode of interrogation cannot provide subsecond response times. Rhodium deposition on the microsensors is much more difficult than that of gold. We were able to attain crack-free Rh deposits by adaptation of pulsed electroplating techniques. At excitation voltages of >2 V dc, the Rh sensors respond to moisture with 10 ↔ 90% rise and fall times of 30−50 ms. These are the fastest microfabricated water vapor sensors reported to date. We demonstrate applications as a breath monitor. Such sensors should also be of utility in atmospheric eddy measurements. Short-term repeatability is better than 0.6% RSD (n = 7). |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/ac0355451 |