Experimental investigation of mass efficiency curve for alpha radioactivity counting using a gas-proportional detector
Gross α counting of evaporated water residues offers a simple method for screening α radioactivity in water for both public health and emergency purposes. The evaporation process for water has been improved by using a combination of roughening of the surface of counting planchettes, two-stage evapor...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2004-06, Vol.60 (6), p.879-886 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Gross
α counting of evaporated water residues offers a simple method for screening
α radioactivity in water for both public health and emergency purposes. The evaporation process for water has been improved by using a combination of roughening of the surface of counting planchettes, two-stage evaporation, and temperature-controlled block heating. The efficiency of the gas-proportional detector for
α-particle detection in water residues was studied as a function of sample mass–thickness in the range between 0.1 and
13
mg
cm
−2
. The effect of
α energy on the efficiency, as well as moisture absorption on the samples, were studied using
230
Th
,
238
U
,
239
Pu
,
241
Am
, and
244
Cm
radionuclides. Also,
α-to-
β crosstalk was investigated as a function of sample mass for
230
Th
,
239
Pu
,
210
Po
,
241
Am
, and
244
Cm
. The improved method can also be applied for gross
α detection in biological fluids. |
---|---|
ISSN: | 0969-8043 1872-9800 |