Validation of a gas chromatographic method for determination of fatty alcohols in 10 mg film-coated tablets of policosanol
A gas chromatographic method using a packed column and 1-eicosanol as an internal standard was validated for the determination of the fatty alcohols that compose policosanol in 10 mg film-coated tablets. The alcohols were determined as trimethylsilyl derivatives, prepared with N-methyl-N-trimethylsi...
Gespeichert in:
Veröffentlicht in: | Journal of AOAC International 2002-05, Vol.85 (3), p.563-566 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A gas chromatographic method using a packed column and 1-eicosanol as an internal standard was validated for the determination of the fatty alcohols that compose policosanol in 10 mg film-coated tablets. The alcohols were determined as trimethylsilyl derivatives, prepared with N-methyl-N-trimethylsilyltrifluoroacetamide. The method can detect degradation products with high retention times, without interfering with the peaks of the active principle. Good linearity (correlation coefficient = 0.9992) and accuracy (mean recovery -100.27 +/- 1.66%) were proven over a range of 25-200% of the nominal concentration. Within- and between-day precision at the nominal 100% value met the acceptance criteria ( |
---|---|
ISSN: | 1060-3271 1944-7922 |
DOI: | 10.1093/jaoac/85.3.563 |