False-positive results and contamination in nucleic acid amplification assays: Suggestions for a prevent and destroy strategy

Contamination of samples with DNA is still a major problem in microbiology laboratories, despite the wide acceptance of PCR and other amplification techniques for the detection of frequently low amounts of target DNA. This review focuses on the implications of contamination in the diagnosis and rese...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of clinical microbiology & infectious diseases 2004-04, Vol.23 (4), p.289-299
Hauptverfasser: BORST, A, BOX, A. T. A, FLUIT, A. C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Contamination of samples with DNA is still a major problem in microbiology laboratories, despite the wide acceptance of PCR and other amplification techniques for the detection of frequently low amounts of target DNA. This review focuses on the implications of contamination in the diagnosis and research of infectious diseases, possible sources of contaminants, strategies for prevention and destruction, and quality control. Contamination of samples in diagnostic PCR can have far-reaching consequences for patients, as illustrated by several examples in this review. Furthermore, it appears that the (sometimes very unexpected) sources of contaminants are diverse (including water, reagents, disposables, sample carry over, and amplicon), and contaminants can also be introduced by unrelated activities in neighboring laboratories. Therefore, lack of communication between researchers using the same laboratory space can be considered a risk factor. Only a very limited number of multicenter quality control studies have been published so far, but these showed false-positive rates of 9-57%. The overall conclusion is that although nucleic acid amplification assays are basically useful both in research and in the clinic, their accuracy depends on awareness of risk factors and the proper use of procedures for the prevention of nucleic acid contamination. The discussion of prevention and destruction strategies included in this review may serve as a guide to help improve laboratory practices and reduce the number of false-positive amplification results.
ISSN:0934-9723
1435-4373