Measurement of nanosecond time‐resolved fluorescence with a directly gated interline CCD camera

Summary CCD cameras coupled optically to gated image intensifiers have been used for fast time‐resolved measurements for some years. Image intensifiers have disadvantages, however, and for some applications it would be better if the image sensor could be gated directly at high speed. Control of the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of microscopy (Oxford) 2002-06, Vol.206 (3), p.233-238
Hauptverfasser: Mitchell, A. C., Wall, J. E., Murray, J. G., Morgan, C. G.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary CCD cameras coupled optically to gated image intensifiers have been used for fast time‐resolved measurements for some years. Image intensifiers have disadvantages, however, and for some applications it would be better if the image sensor could be gated directly at high speed. Control of the ‘charge drain’ function on an interline‐transfer CCD allows the sensor to be switched rapidly from an insensitive state. The temporal and spatial properties of the charge drain are explored in the present paper and it is shown that nanosecond time resolution with acceptable spatial uniformity can be achieved for a small commercial sensor. A fluorescence lifetime imaging system is demonstrated, based on a repetitively pulsed laser excitation source synchronized to the CCD control circuitry via a programmable delay unit.
ISSN:0022-2720
1365-2818
DOI:10.1046/j.1365-2818.2002.01030.x