Automatic tuning and matching of a small multifrequency saddle coil at 4.7 T

A new circuit design for automatically tuning and matching a saddle coil for small animal imaging is presented. This design allows working at 1H, 19F, and 3He resonance frequencies in a 4.7 T spectrometer. It is based on a balanced circuit with commercial variable capacity diodes controlled by a com...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2004-04, Vol.51 (4), p.869-873
Hauptverfasser: Pérez de Alejo, Rigoberto, Garrido, Carlos, Villa, Palmira, Rodriguez, Ignacio, Vaquero, Juan José, Ruiz-Cabello, Jesús, Cortijo, Manuel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new circuit design for automatically tuning and matching a saddle coil for small animal imaging is presented. This design allows working at 1H, 19F, and 3He resonance frequencies in a 4.7 T spectrometer. It is based on a balanced circuit with commercial variable capacity diodes controlled by a computer using digital potentiometers. The change between two different frequencies can be accurately performed in a few seconds. System Q is compared, between 140–210 MHz, to the same coil tuned and matched with high Q variable capacitors. Differences lower than 5% were found with a loaded coil. The proposed design has initially been evaluated in 19F and 1H NMR images acquired with a five‐tube phantom. An application is also shown for the acquisition of 3He, 19F, and 1H lung images in a control rat. Magn Reson Med 51:869–873, 2004. © 2004 Wiley‐Liss, Inc.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.20045