Application of radiofrequency energy in surgical and interventional procedures: are there interactions with ICDs?

During surgical and interventional procedures, interference may occur between ICDs and electrical cautery or with the application of RF energy. This may lead to the false induction of ICD therapies or could even result in device malfunction, which represents a potential perioperative hazard for the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2004-03, Vol.27 (3), p.293-298
Hauptverfasser: Fiek, Michael, Dorwarth, Uwe, Durchlaub, Ilka, Janko, Sabine, Von Bary, Christian, Steinbeck, Gerhard, Hoffmann, Ellen
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:During surgical and interventional procedures, interference may occur between ICDs and electrical cautery or with the application of RF energy. This may lead to the false induction of ICD therapies or could even result in device malfunction, which represents a potential perioperative hazard for the patient. This study analyzed the intraoperative interactions in 45 consecutive ICD patients in reference to different surgical and interventional procedures. A total of 33 surgical operations (general surgery [n = 14], urologic [n = 5], abdominal [n = 10], gynecological [n = 2], thoracic [n = 1], neurosurgical [n = 1]) and 12 interventional therapies (RF catheter ablation [n = 10], endoscopic papillotomy [n = 2]) were performed. The ICD devices were all located in left pectoral position and consisted of 25 single and 20 dual chamber defibrillators. During the procedure, tachyarrhythmia detection (VF 296 +/- 20 ms, VT 376 +/- 49 ms) of the devices was maintained active (monitoring mode), only ICD therapies were inactivated. The indifferent electrode of the electrical cauter/RF generator was placed in standard positions (right/left mid-femoral position [n = 27/8], thoracic spine area [n = 10]). After the procedure, the ICD memory was checked for detections and for changes in the programming. There was no oversensing, reprogramming, or damage of any defibrillator caused by RF energy. Despite the lack of undesired interactions, ICDs should be inactivated preoperatively to assure maximum patient safety. However, should inactivation not be possible, or the achievement uncertain, electromagnetic interference is highly unlikely.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.2004.00430.x