Nanoflow Gradient Generator for Capillary High-Performance Liquid Chromatography

A novel method of generating a nanoflow gradient elution for a capillary high-performance liquid chromatography (HPLC) system has been developed. An important feature of this system is that any gradient (GR) profile generated by a conventional microflow GR pump can be asymptotically traced and conve...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2004-03, Vol.76 (5), p.1524-1528
Hauptverfasser: Deguchi, Kisaburo, Ito, Shinya, Yoshioka, Shinji, Ogata, Izumi, Takeda, Akihiro
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A novel method of generating a nanoflow gradient elution for a capillary high-performance liquid chromatography (HPLC) system has been developed. An important feature of this system is that any gradient (GR) profile generated by a conventional microflow GR pump can be asymptotically traced and converted as a corresponding nanoflow GR profile simply by using a 10-port switching valve with two injection loops installed. Consequently, it has been called an “asymptotic trace 10-port valve” (AT10PV) nanoflow GR generator. Performance of the AT10PV nanoflow GR generator was tested in the range of flow rates from 50 to 500 nL/min. The test demonstrated that the AT10PV nanoflow GR generator can asymptotically trace the original gradient profile with good reproducibility. A capillary HPLC system using the AT10PV nanoflow GR generator provides reasonably good repeatability of peak retention times on the chromatogram of the tryptic digest of a BSA sample, RSD of less than 0.3% at a flow rate of 200 nL/min. It also enables sequential running of a series of sample injections in the same manner as conventional analysis at microflow rates.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac0350312