Use of Sample Hematocrit Value to Correct Blood Tacrolimus Concentration Derived by Microparticle Enzyme Immunoassay
The quality of microparticle enzyme immunoassay (MEIA) for blood tacrolimus is guaranteed in samples with hematocrit (Ht) values of 25 to 45%. Because MEIA provides inaccurate blood tacrolimus concentrations in samples with Ht out of this range (i.e. 45%), correction of the calibration is required f...
Gespeichert in:
Veröffentlicht in: | Biological & Pharmaceutical Bulletin 2008/06/01, Vol.31(6), pp.1250-1253 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The quality of microparticle enzyme immunoassay (MEIA) for blood tacrolimus is guaranteed in samples with hematocrit (Ht) values of 25 to 45%. Because MEIA provides inaccurate blood tacrolimus concentrations in samples with Ht out of this range (i.e. 45%), correction of the calibration is required for therapeutic drug monitoring. The authors demonstrated previously that overestimated MEIA tacrolimus concentration could be corrected by modified, calibrated MEIA (cMEIA) using the original calibrator. Here, an equation was established to more easily derive a corrected tacrolimus concentration by calculation (MEIAcalc) using the Ht of each sample. The tacrolimus concentrations of 99 whole-blood samples with low Ht ( |
---|---|