Comparison of flat-panel detector and image-intensifier detector for cone-beam CT

We evaluated a flat-panel detector (FPD) (scintillator screen and a-Si photo-sensor array) for use in a cone-beam computed tomography (CT) detector and compared it with an image-intensifier detector (IID). The FPD cone-beam CT system has a higher spatial resolution than the IID system. At equal pixe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Computerized medical imaging and graphics 2002-05, Vol.26 (3), p.153-158
Hauptverfasser: Baba, Rika, Konno, Yasutaka, Ueda, Ken, Ikeda, Shigeyuki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We evaluated a flat-panel detector (FPD) (scintillator screen and a-Si photo-sensor array) for use in a cone-beam computed tomography (CT) detector and compared it with an image-intensifier detector (IID). The FPD cone-beam CT system has a higher spatial resolution than the IID system. At equal pixel sizes, the standard deviation of noise intensity of the FPD system is equal to that of the IID system. However, the circuit noise of the FPD must be reduced, especially at low doses. Our evaluations show that the FPD system has a strong potential for use as a cone-beam CT detector because of high-spatial resolution.
ISSN:0895-6111
1879-0771
DOI:10.1016/S0895-6111(02)00008-3