Endotension is Influenced by Aneurysm Volume: Experimental Findings
Purpose: To investigate in an in vitro model whether and to what extent pressure is influenced by aneurysm size. Methods: Latex aneurysms of 3 different volumes (24, 30, and 81 mL) were inserted into an in vitro circulation model. The systemic mean pressure (SPmean) was varied from 50 to 120 mmHg. T...
Gespeichert in:
Veröffentlicht in: | Journal of endovascular therapy 2003-12, Vol.10 (6), p.1091-1096 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
To investigate in an in vitro model whether and to what extent pressure is influenced by aneurysm size.
Methods:
Latex aneurysms of 3 different volumes (24, 30, and 81 mL) were inserted into an in vitro circulation model. The systemic mean pressure (SPmean) was varied from 50 to 120 mmHg. The aneurysms were excluded using a woven polyethylene graft. Aneurysm sac mean pressure (ASPmean) was measured.
Results:
In the in vitro model, endovascular aneurysm repair created a closed chamber without endoleak but showed a relevant aneurysm sac pressure. At an SPmean of 80 mmHg, the ASPmean was 42.0 ± 0.6 mmHg in the 24-mL aneurysm, 40.5 ± 0.5 mmHg in the 30-mL model, and 19.3 ± 0.5 mmHg in the 81-mL aneurysm (p < 0.05). The ASPmean rose with increasing SPmean and was inversely dependent on the aneurysm volume.
Conclusions:
This in vitro model demonstrated that the sac mean pressure correlated to the systemic pressure and that a greater aneurysm volume reduced aneurysm sac pressure. These data highlight the need for further studies regarding endotension. |
---|---|