Detection of 27 Molecules in a Single Injection Volume by Hadamard Transform Capillary Electrophoresis
A concentration detection limit of 100 fM was achieved for the fluorescein ion by improving the experimental setup used for Hadamard transform capillary electrophoresis. Two argon-ion lasers, a gating laser for sample injection and a probe laser for the excitation of analyte molecules, were employed...
Gespeichert in:
Veröffentlicht in: | Analytical Sciences 2003, Vol.19(12), pp.1659-1661 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A concentration detection limit of 100 fM was achieved for the fluorescein ion by improving the experimental setup used for Hadamard transform capillary electrophoresis. Two argon-ion lasers, a gating laser for sample injection and a probe laser for the excitation of analyte molecules, were employed for the efficient photodegradation of analyte molecules in laser-induced fluorescence detection using an optically gated sample-injection method. In addition, a dichroic mirror, located in the pathway of the probe laser was used to exclude the other lines of the argon-ion laser. Using a Hadamard matrix on the order of 2046, the concentration limit of detection for fluorescein ion was determined to be 100 fM at S/N = 3, in which the average number of molecules in a single injection volume was calculated to be 27. The influences of the output power in both the gating and probe lasers on the sensitivity are also discussed. |
---|---|