Biomechanical investigations of different osteosynthesis procedures in the proximal tibia-introduction of a new measuring system
In this article a new measuring system for biomechanical studies of various osteosynthesis procedures on the tibia is described. The paired tibias are osteotomized and embedded in PMMA in accordance with a randomization protocol. Testing is then carried out under nonaxial loading at 350 N, 600 N and...
Gespeichert in:
Veröffentlicht in: | Biomedizinische Technik 2003-11, Vol.48 (11), p.319-324 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | ger |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this article a new measuring system for biomechanical studies of various osteosynthesis procedures on the tibia is described. The paired tibias are osteotomized and embedded in PMMA in accordance with a randomization protocol. Testing is then carried out under nonaxial loading at 350 N, 600 N and 900 N, uniaxial bending applying a force of 12 Nm, and torsion at a force of 5 Nm in a universal pneumatic testing machine. Ultrasonic sensors pick up movement at the fracture gap under external loading. Testing of the measuring setup has been successful, and relevant results were obtained. |
---|---|
ISSN: | 0013-5585 |