The magnitude of signal errors introduced by ISIS in quantitative 31P MRS
It is well known that the quality of a quantitative 31P MRS measurement relies largely on the performance of the volume selection method, and that image selected in vivo spectroscopy (ISIS) suffers from contaminating signal caused mostly by T1 smearing. However, these signal errors and their magnitu...
Gespeichert in:
Veröffentlicht in: | Magma (New York, N.Y.) N.Y.), 2002-03, Vol.14 (1), p.30-38 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | It is well known that the quality of a quantitative 31P MRS measurement relies largely on the performance of the volume selection method, and that image selected in vivo spectroscopy (ISIS) suffers from contaminating signal caused mostly by T1 smearing. However, these signal errors and their magnitude are seldom addressed in clinical studies. The aim of this study was therefore to investigate the magnitude of signal errors in 31P MRS when using ISIS. The results from the measurements with a homogeneous head phantom are as follows: at low TR/T1 ratios the contamination increases rapidly, especially for small ( |
---|---|