Relationship Between Quantec Measurement and Cobb Angle in Patients with Idiopathic Scoliosis

The Quantec Spinal Imaging System (QSIS) is a raster stereography used to measure three-dimensional trunk images. The Q angle, a coronal plane measurement generated by the Quantec Spinal Imaging System (QSIS), was compared with the Cobb angle in assessment of scoliosis curve magnitude. One hundred f...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of pediatric orthopaedics 2000-07, Vol.20 (4), p.512-516
Hauptverfasser: Thometz, J G, Lamdan, R, Liu, X C, Lyon, R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Quantec Spinal Imaging System (QSIS) is a raster stereography used to measure three-dimensional trunk images. The Q angle, a coronal plane measurement generated by the Quantec Spinal Imaging System (QSIS), was compared with the Cobb angle in assessment of scoliosis curve magnitude. One hundred forty-nine patients with idiopathic scoliosis were evaluated using both the Quantec system and plane radiographs. The Cobb and Q angles demonstrated significant correlation in the thoracic region (r = 0.65, p < 0.05), lumbar region (r = 0.63, p < 0.05), and in the thoracolumbar region (r = 0.70, p < 0.05). The difference between the Q and Cobb angles was small when the Cobb angle was
ISSN:0271-6798
1539-2570
DOI:10.1097/01241398-200007000-00017