Only carbon dioxide absorbents free of both NaOH and KOH do not generate compound A during in vitro closed-system sevoflurane: Evaluation of five absorbents

Insufficient data exist on the production of compound A during closed-system sevoflurane administration with newer carbon dioxide absorbents. A modified PhysioFlex apparatus (Dräger, Lübeck, Germany) was connected to an artificial test lung (inflow at the top of the bellow approximately/= 160 ml/min...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anesthesiology (Philadelphia) 2001-09, Vol.95 (3), p.750-755
Hauptverfasser: VERSICHELEN, Linda F. M, BOUCHE, Marie-Paule L. A, ROLLY, Georges, VAN BOCXLAER, Jan F. P, STRUYS, Michel M. R. F, DE LEENHEER, André P, MORTIER, Eric P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Insufficient data exist on the production of compound A during closed-system sevoflurane administration with newer carbon dioxide absorbents. A modified PhysioFlex apparatus (Dräger, Lübeck, Germany) was connected to an artificial test lung (inflow at the top of the bellow approximately/= 160 ml/min CO2; outflow at the Y piece of the lung model approximately/= 200 ml/min, simulating oxygen consumption). Ventilation was set to obtain an end-tidal carbon dioxide partial pressure of approximately 40 mmHg. Various fresh carbon dioxide absorbents were used: Sodasorb (n = 6), Sofnolime (n = 6), and potassium hydroxide (KOH)-free Sodasorb (n = 7), Amsorb (n = 7), and lithium hydroxide (n = 7). After baseline analysis, liquid sevoflurane was injected into the circuit by syringe pump to obtain 2.1% end-tidal concentration for 240 min. At baseline and at regular intervals thereafter, end-tidal carbon dioxide partial pressure, end-tidal sevoflurane concentration, and canister inflow (T degrees(in)) and canister outflow (T degrees(out)) temperatures were measured. To measure compound Ainsp concentration in the inspired gas of the breathing circuit, 2-ml gas samples were taken and analyzed by capillary gas chromatography plus mass spectrometry. The median (minimum-maximum) highest compound Ainsp concentrations over the entire period were, in decreasing order: 38.3 (28.4-44.2)* (Sofnolime), 30.1 (23.9-43.7) (KOH-free Sodasorb), 23.3 (20.0-29.2) (Sodasorb), 1.6 (1.3-2.1)* (lithium hydroxide), and 1.3 (1.1-1.8)* (Amsorb) parts per million (*P < 0.01 vs. Sodasorb). After reaching their peak concentration, a decrease for Sofnolime, KOH-free Sodasorb, and Sodasorb until 240 min was found. The median (minimum-maximum) highest values for T degrees(out) were 39 (38-40), 40 (39-42), 41 (40-42), 46 (44-48)*, and 39 (38-41) degrees C (*P < 0.01 vs. Sodasorb), respectively. With KOH-free (but sodium hydroxide [NaOH]-containing) soda limes even higher compound A concentrations are recorded than with standard Sodasorb. Only by eliminating KOH as well as NaOH from the absorbent (Amsorb and lithium hydroxide) is no compound A produced.
ISSN:0003-3022
1528-1175
DOI:10.1097/00000542-200109000-00030