Compound A production from sevoflurane is not less when KOH-free absorbent is used in a closed-circuit lung model system

In an in vitro study, less compound A was formed when a KOH-free carbon dioxide absorbent was used. To confirm this observation we used a lung model in which carbon dioxide was fed in at 160 ml min−1 and sampling gas was taken out for analysis at 200 ml min−1; ventilation aimed for a Pe′CO2 of 5.4 k...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 2001-03, Vol.86 (3), p.345-348
Hauptverfasser: Versichelen, L, Bouche, M.P., Struys, M, Van Bocxlaer, J, Mortier, E, de Leenheer, A.P., Rolly, G
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In an in vitro study, less compound A was formed when a KOH-free carbon dioxide absorbent was used. To confirm this observation we used a lung model in which carbon dioxide was fed in at 160 ml min−1 and sampling gas was taken out for analysis at 200 ml min−1; ventilation aimed for a Pe′CO2 of 5.4 kPa. The soda lime canister temperatures in the inflow and outflow ports (Tin and Tout) were recorded. In six runs of 240 min each, a standard soda lime, Sodasorb (Grace, Epernon, France) was used and in eight runs KOH-free Sofnolime (Molecular Products, Thaxted, UK) was used. Liquid sevoflurane was injected using a syringe pump to obtain 2.1% e′. Compound A was measured by capillary gas chromatography combined with mass spectrometry. Median (range) compound Ainsp increased to a maximum of 22.7 (7.9) ppm for Sodasorb and 33.1 (20) for Sofnolime at 60 min and decreased thereafter; the difference between groups was significant (P
ISSN:0007-0912
1471-6771
DOI:10.1093/bja/86.3.345