Glucose infusion test: A new screening test for vascular access recirculation
Glucose infusion test: A new screening test for vascular access recirculation. Vascular access recirculation is an important cause of diminished dialysis efficiency. We propose a new screening test based on glucose infusion as a tracer for recirculation. The glucose infusion test (GIT) protocol comp...
Gespeichert in:
Veröffentlicht in: | Kidney international 2000-05, Vol.57 (5), p.2123-2128 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Glucose infusion test: A new screening test for vascular access recirculation.
Vascular access recirculation is an important cause of diminished dialysis efficiency. We propose a new screening test based on glucose infusion as a tracer for recirculation.
The glucose infusion test (GIT) protocol comprises a basal blood sample (A) from the arterial port, a 5 mL bolus of 20% glucose into the venous chamber (time 0), followed by a second sample (B) in four seconds (from 13 to 17 s with QB 300 mL/min) from the same port. The blood glucose level is determined at the bedside on A and B with a reflectance photometer (CV 1.8%). Interpretation of the test is straightforward: If B = A, there is no recirculation, whereas if B > A, recirculation can be calculated from the regression equation: 0.046 × (B - A) + 0.07, obtained from in vitro tests reproducing artificial recirculation at 0, 5, and 10%. To validate this new method in vivo, we compared GIT and the urea test on 39 hemodialysis patients, obtaining a good correlation (r = 0.93). The two tests were considered positive (recirculation present) when the lower 95% confidence intervals were more than zero.
Our patients were divided into two groups: those with (22 out of 39, mean recirculation 11.8%) or without recirculation (17 out of 39, mean 0.06%). The urea test did not recognize 7 out of 22 patients because they had a small recirculation below the urea test limit of detection.
GIT was more sensitive (detection limit 0.3%), simpler, and immediate in showing the results than the urea test. It is an accurate and low-cost technique for screening and follow-up of vascular access in a dialysis unit. |
---|---|