Experimental verification of the properties of ion-selective electrodes for low concentration determination

The influence of several ion-selective electrode properties on electrode response and selectivity at low concentration levels has been investigated experimentally. The properties investigated were the composition of inner electrode solution, the composition of the membrane, the presence of interferi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Fresenius' journal of analytical chemistry 2001-06, Vol.370 (4), p.367-370
Hauptverfasser: SOKALSKI, Tomasz, BEDLECHOWICZ, Iwona, MAJ-ZURAWSKA, Magdalena, HULANICKI, Adam
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The influence of several ion-selective electrode properties on electrode response and selectivity at low concentration levels has been investigated experimentally. The properties investigated were the composition of inner electrode solution, the composition of the membrane, the presence of interfering ions in the sample, and the thickness of diffusion layer in the sample solution. All the results obtained confirmed theoretical considerations.
ISSN:0937-0633
1432-1130
DOI:10.1007/s002160100799